InterTest Microprocessor Controlled Gross/Fine Leak Pressurization System With LCD Display
InterTest 1014-CII Systems are available in many configurations including 6", 9", or 12" diameter chambers.

Easy to Operate - Microprocessor Control System provides interactive communication with the operator and fast, simple and flexible programming.
Large Liquid Crystal Display and Touch-Key Entry 10" x 2 1/2" - Display constantly gives current status for each chamber and provides instant review of all settings. Simply press the start key to initiate a cycle. Two programs are stored for each chamber... one for gross and one for fine. Easy to Maintain and Calibrate
Real Time Diagnostics - Continuously self-monitored system displays error messages in plain English for quick, effective troubleshooting.
Highly Accessible Design - Easily removable panels, strategically placed components and tough, flexible Parflex tubing permit easy access for maintenance and service. Manual Control Panel The turn of a keyswitch instantly sends the status of all vacuum, pressure and fluid level readings to the display for easy calibration. Additionally, all system valves and pumps may be individually controlled.
High Reliability, Integrity and Productivity
Full Password Protection - System may be programmed to permit test parameter changes by authorized personnel only. Exceeds MIL-STD-883 and High-Rel Specifications Designed to meet or exceed Mil STD 883, Method 1014, Mil STD 750, Method 1071, Mil STD 202, Method 112 and other High-Rel commercial equivalents for Gross and Fine Leak Test conditions.
Power Failure Recovery - In the event of a power failure all values are stored into battery backed memory, and the computer maintains control of the cycle. Upon restoration, all test parameters are verified to ensure that the cycle remains within compliance, then the cycle continues, uninterrupted. Automatic Start Yields 33% Throughput Increase Either or both chambers may be loaded at the end of a shift/day and programmed to automatically complete preconditioning as the next shift/day begins. "Fluid Level Restore/Helium Hold" Modes "Fluid Level Restore" maintains chamber fluid level above tubes as each is removed for testing. "Helium Hold" Mode maintains pressure setpoint which allows operators convenient flexibility between preconditioning and helium leak testing.
Totally Independent Chamber Operation - Process chambers are completely independently programmable and function separately to allow flexibility for balancing workloads.
Fluorocarbon Savings - In the ready mode, fluid detected at the bottom of a chamber is automatically drained to storage. This allows an operator to empty remaining fluid from buckets into the chamber, thus reducing handling losses. The use of redundant vacuum valves prevents fluorocarbon ingestion by the pump. A 3-way vent valve properly routes fluorocarbon laced nitrogen through the storage tank for recondensing and routes helium directly to atmosphere disallowing any fluorocarbon evaporation. The system constantly monitors for unwanted fluorocarbon presence during vacuum cycles. In this case, the cycle is aborted and the operator is notified, unlike competitors' systems.
Available Options
Printer to provide test certification for full traceability
Helium only
Single Chamber
6", 9" or 12" Diameter chambers (any depth to 30")
Parts Handlers
Filler Blocks
Service Contracts
Specifications
Activation Chambers (Standard): 6" ID x 28"L (3.6 gallon capacity)
9" ID x 28"L (8.1 gallon capacity)
12" ID x 28"L (13.7 gallon capacity)
Vacuum System: 7 CFM Industrial direct drive pump
Fluorocarbon System: Filtration to 0.45 Micron, individual rotary gear pumps for fill and drain
Dimensions/Weight: 48"W x 32"D x 55"H, 800 lbs.
Gas Supply Requirements: Regulated N2 or CDA, and/or He @ 95 PSIG
Power: 110-120 VAC, 60 Hz, 30A (Domestic) or 220-240 VAC, 50 Hz, 30A (Intntl)
Special Features: Audible announcement of cycle completion, diagnostic display messages on LCD should an error occur, battery backed memory retention of system settings, automatic fluid level restore for stick mode, external fluid fill and drain.
InterTest Helium Preconditioning System for Hermetic Seal Testing.
The 1014-CII/He Helium Preconditioning System with dual front load rectangular chambers each provide spacious 14.8"W x 7.8"H x 18"L capacity ideal for applications where large component trays or boats are utilized.

Microprocessor Control
Large LCD Display and Touch-Key Entry
Full Password Protection
Power Failure Recovery
Automatic Start Yields 33% Throughput Increase
Totally Independent Chamber Operation
Helium "Hold" Mode
Exceeds MIL-STD-883 and High-REL Specifications
InterTest Bubble Detection Systems
The InerTest 1014-CBL Gross Leak Bubble Detection System features built-in filtration, lighting and magnification. These self-contained units are available in any size to meet your test space requirements.
Features
Exceeds Requirements of MIL-STD 883, 750 and 202.
Built-in 3x Magnifying Lens
Dual Illumination Lamps each exceed 18,000 Ft-candles
Digital temperature controller using a precision platinum RTD Sensor displays tank temperature to 0.1 degrees C with easy front panel access for calibration and setpoint adjustments.
Fluorocarbon Saver Condenser Ring reduces costly evaporation particularly when used in conjunction with InterTest Coolant Circulation System (Model 1014-CBC).
Built-in 0.45 Micron Filtration System may be initiated with fluid at operating temperature. After filtration is complete, the fluid returns to operating temperature automatically. (Ideal for operator lunch breaks and between shifts.)
Dual Horizontal Heating Elements, each capable of bringing fluid to 125 degrees C. (If one element fails, the other enables operator to continue to test).
All Stainless Steel construction with 1/4" thick Pyrex glass
Microprocessor Control
Over Temperature protection circuit
Hundreds of field-tested units
Larger capacity units available (Models 1014-CBL/X & 1014-CBL/S)
Spec
Dimensions/Weight: 22"W x 17"D x 12"H, 65 lbs
Basket Dimensions: 10"W x 4"
Front to Back
Power: 115 VAC, 10A, 50/60 HZ
Electrical: Circuit Breaker Protected (no fuses)
Fluid Capacity: 2.12 U.S. Gallons
Pump/Filter Capacity: 4 GPM, 0.45 Micron
Test Timers: 30/60 Seconds Selectable
Refrigerated Centrifuge System
InterTest's 2001-E Centrifuge represents the latest advancements in safety, performance and ergonomic design for constant acceleration testing of microelectronic components. The 2001-E meets or exceeds MIL-STD-883, Method 2001, Condition A-E; MIL-STD-750, and other HIGH-REL requirements.

User-Friendly Design - Exceptionally quiet operation Unique design keeps operating noise levels at less than 70 db. Easy-open chamber Non-vacuum design eliminates the need for bulky chamber door.
Easy to Operate and Maintain - Built-in Diagnostics Continuously self-monitored system displays error messages for 12 condition parameters. Low Maintenance Design Direct drive eliminated the need for belt replacement and increases spindle and bearing life. Fixture Compatibility Full compatibility with most standard industry fixtures eliminates the need for large fixture inventories. Easy Speed Verification Easy access external jacks ensure conformance to speed parameters.
High Reliability and Integrity - Exceeds MIL-STD-883 and High-Rel Specifications The system is designed to meet or exceed MIL-STD-883, Method 2001, Conditions A through E, MIL-STD-750, Method 2006, and other HIGH-REL commercial equivalents. Built-in Refrigeration System Refrigeration system with full, 5-year warranty eliminates vacuum related failures. System provides highly accurate temperature control and cool operation for maximum bearing life. Patented Direct-Drive Provides precise spin-speed control while eliminating stress and maintenance problems associated with belt-drive systems
Safe Operation - Safety-Plus Chamber Design The Safety-Plus chamber is designed with a stainless steel chamber wall, cushioned by a layer of insulating foam and surrounded by a full 1/2" thick carbon steel containment ring. The containment ring rotates to absorb energy in the event that load separation penetrates the first two protective layers, thereby keeping the centrifuge in place on the floor. In addition, the fully interlocked door is constructed with over 1 1/2" of lightweight, energy absorbing material backed by 1/4" carbon steel to safely contain any minimal upward forces that might occur during load separation. Out-of-Balance Protection Advanced, non-contact piezoelectric crystal immediately detects when out-of-balance condition exists. Interlocks and Alarms Includes door closure interlocks and alarms for overspeed, overtemp, out-of-balance, entry error and other faults.
Specification
Controls/Displays: G-Force: 0-32,700 G
Spin Speed: 0-19,940 rpm
Cycle Time: 0-99 min/sec
Test Radius: 2.0" to 5.0"
Motor: Direct Drive
Acceleration: 0-16,000 rpm
<90 sec (E-70)
<120 sec (E-50)
Deceleration: 16,000-0 rpm
<90 sec (E-70)
<120 sec (E-50)
Overall Dimensions: 30"W x 39"D x 48"H
Weight (net lbs): 800
Power: 208 or 230 VAC, 60 Hz, 40 A
220 or 240 VAC, 50 Hz, 40 A